Sterilization monitoring

Sterilization is checked three ways: mechanical readings every cycle, chemical indicators on and inside every pack, and biological spore tests at least weekly and with every implant load. Steam and chemical vapor use Geobacillus stearothermophilus; dry heat uses Bacillus atrophaeus. A positive spore test takes the unit out of service for a retest, and a second positive means recall and repair.

High-yield facts · 7

  1. Monitor mechanically (time, temperature, pressure) every cycle; use chemical indicators on and inside every package.S10S7
  2. Spore testing (biological indicators) at least weekly and with every load containing an implantable device.S10S7
  3. Steam and unsaturated chemical vapor: Geobacillus stearothermophilus spores. Dry heat: Bacillus atrophaeus spores.S10
  4. Prevacuum steam sterilizers also need a daily air-removal (Bowie-Dick type) test before the first load.S9
  5. External chemical indicators (e.g., tape) show exposure only; internal integrating indicators react to time, temperature and steam.S10
  6. A positive spore test: take the sterilizer out of service, check procedures, retest; if positive again, recall items processed since the last negative test and repair the unit.S10
  7. Keep sterilization records for the time required by state law.S10